Automatic chemistry analyser
Cormay ACCENT-MC240
Cormay ACCENT MC240 is a benchtop automatic chemistry analyzer for professional in-vitro laboratory use. It combines 80 reusable cuvettes, an independent mixing bar and a diffraction-grating optical system with 12 wavelengths. The manufacturer specifies 240 tests/hour in the base mode; up to 400 tests/hour applies to a configuration with optional ISE. A refrigerated sample/reagent compartment and built-in barcode reader support sample workflow. Before selecting a configuration, check reagent and consumable compatibility together with installation requirements.
Catalog code: F10042Source item code: F10042Brand: Cormay
1,168,500 UAH
Recorded on 28.10.2026 · Source product page ↗
FORMED VIP price and availability are unconfirmed. Call to clarify current terms and included equipment.
About the model
Cormay ACCENT MC240 is a benchtop automatic chemistry analyzer for professional in-vitro laboratory use. It combines 80 reusable cuvettes, an independent mixing bar and a diffraction-grating optical system with 12 wavelengths. The manufacturer specifies 240 tests/hour in the base mode; up to 400 tests/hour applies to a configuration with optional ISE. A refrigerated sample/reagent compartment and built-in barcode reader support sample workflow. Before selecting a configuration, check reagent and consumable compatibility together with installation requirements.
ISE is an option for measuring K+, Na+ and Cl−; its higher throughput is not a base-supply specification. The manufacturer describes a closed system with three open channels. Confirm available methods and specific reagent compatibility against the current instructions rather than assuming a universal assay menu.
Supplied contents: the manufacturer distinguishes B05-A0001 with a barcode reader and clot detector from B05-A0002 with an additional ISE module. The exact factory catalogue number and complete contents of the formed.ua offer have not been confirmed. Do not infer that ISE, reagents, controls, calibrators, a computer or accessories are included from photographs or lists of compatible components.
Model specifications were checked against Cormay’s public product page and technical leaflet CC_ACC240_LL_05_2026_ENG. Price and photographs remain sourced from formed.ua; factory specifications do not establish the contents of the particular reseller offer.
Operating atmospheric pressure: 80–106 kPa; Storage pressure: 50–106 kPa; Cuvette arrangement: 8 segments of 10 reusable cuvettes.
Price and availability were checked only on formed.ua, not confirmed for FORMED VIP. For trained laboratory personnel. Verify supplied contents, reagent compatibility and use in the instructions; the photo does not establish inclusion of additional items.
Specifications
- Base-mode throughput
- 240 tests/h
- Throughput with optional ISE
- Up to 400 tests/h, calculated value
- Cuvettes
- 80 reusable
- Sample volume
- 2–35 µL; step 0.1 µL
- Reagent volume
- 10–200 µL; step 0.5 µL
- Reaction volume
- 100–360 µL
- Compartment cooling
- 2–12 °C
- Positions on manufacturer page
- 50 reagent; up to 100 sample
- Total tray capacity in technical leaflet
- Up to 150 total for samples and reagents
- Optical system
- Concave diffraction grating, 12 wavelengths
- Wavelengths
- 340 / 380 / 412 / 450 / 505 / 546 / 570 / 605 / 660 / 700 / 740 / 800 nm
- Photometric range
- 0–3.5 Abs
- Light source
- Halogen-tungsten lamp
- Water consumption in technical leaflet
- ≤6.5 L/h
- Sample carry-over
- ≤0.05%
- Mixing
- Independent mixing bar
- Barcodes
- Built-in reader; bidirectional LIS
- Dimensions — length × depth × height
- ≤860 × 660 × 550 mm
- Weight
- 115 ± 15 kg
- Rated apparent power consumption
- ≤1300 VA
- Operating temperature
- 15–30 °C
- Operating relative humidity
- 35–85% (Non-condensing)
- Storage temperature
- 0–40 °C
- Power supply
- 220–240 V~ 50 Hz; 220/230 V~ 60 Hz
- Interfaces
- RS232; network; USB / parallel; VGA / DVI
- Workstation operating system in leaflet
- Windows 10 (64-bit)
- Analysis methods
- End-point, fixed-time, kinetic; ISE optional
- Urgent samples
- STAT priority
- Operating altitude
- −400–2000 m
- Storage humidity
- 30–85%, non-condensing
- Probe functions
- Liquid level, horizontal/vertical obstruction, clot detection and level tracking
- Probe cleaning
- Automatic interior and exterior washing
- Cuvette cleaning
- Station with pre-warmed detergent and water
- Calibration
- K-factor; linear two-/multi-point; non-linear: Logit-Log 4P/5P, spline, exponential, polynomial, parabola, Logit-Log 3P, broken line
- Quality control
- Westgard multirules; cumulative sum; twin-plot; Levey–Jennings
- Operating atmospheric pressure
- 80–106 kPa
- Storage pressure
- 50–106 kPa
- Cuvette arrangement
- 8 segments of 10 reusable cuvettes